Temperature Mapping Procedures

  1. Preparation:
  • Identify the equipment or area to be mapped (e.g., fridges, freezers, warehouses).
  • Ensure the equipment is empty or use dummy products to simulate normal conditions
  1. Sensor Placement:
  • Distribute calibrated sensors (data loggers) evenly throughout the space.
  • Typically, a 9 or 10-point mapping is used, with sensors placed in each corner at the top and bottom, and one in the centre
  1. Data Collection:
  • Leave the sensors in place for a set period, ranging from 24 hours to several days, depending on the size of the area
  • Record temperature data continuously to capture fluctuations and identify any hot or cold spots
  1. Analysis:
  • Analyse the collected data to determine temperature variations within the space.
  • Identify areas that may require adjustments or additional monitoring

 

  • Reporting and Action:
  • Document the findings and create a report detailing the temperature profile of the equipment.
  • Make necessary adjustments to the equipment or sensor placement based on the results

 

 

Calibration

  1. Initial Calibration:
  • Perform an initial calibration using a traceable thermometer to ensure accuracy
  • For critical applications, consider a UKAS calibration, which follows ISO 17025 standards
  1. Regular Calibration:
  • Schedule regular calibrations, typically once a year, to maintain accuracy
  • Recalibrate whenever significant modifications are made to the equipment or if there are changes in regulatory requirements
  1. Documentation:
  • Keep detailed records of all calibration activities, including the methods used and the results obtained